Difficulty: Medium
Average Score: 94%
Testbank
When the nurse monitors the water-sealed drainage system, which finding suggests the system is working properly?
Correct answer
A
Rationale
Fluid rises and falls with respirations.
This finding indicates that the water-sealed drainage system is functioning correctly, as it demonstrates the presence of negative pressure in the pleural space during inhalation and exhalation, allowing for proper lung expansion and drainage of air or fluid.
B: Level of fluid is lowered in suction chamber. This could indicate a malfunction or obstruction in the system rather than proper function, as it does not reflect the dynamic changes that occur with respiration.
C: Fluid is bubbling vigorously. While some bubbling may be normal, vigorous bubbling often suggests a leak in the system, indicating that it is not functioning properly.
D: Fluid appears white and frothy. This appearance may suggest the presence of fluid or blood in the drainage, which is not a sign of proper functioning but rather a potential complication needing further assessment.
This finding indicates that the water-sealed drainage system is functioning correctly, as it demonstrates the presence of negative pressure in the pleural space during inhalation and exhalation, allowing for proper lung expansion and drainage of air or fluid.
B: Level of fluid is lowered in suction chamber. This could indicate a malfunction or obstruction in the system rather than proper function, as it does not reflect the dynamic changes that occur with respiration.
C: Fluid is bubbling vigorously. While some bubbling may be normal, vigorous bubbling often suggests a leak in the system, indicating that it is not functioning properly.
D: Fluid appears white and frothy. This appearance may suggest the presence of fluid or blood in the drainage, which is not a sign of proper functioning but rather a potential complication needing further assessment.